Product Name JingQi FP6410 Flame Photometer Yoke FP6410 FP Series Flame PhotoMeter
Product Price $1185.00 In coming
Details Details
Model comparison
Channels Dual Dual
Elements Na K Na K Li Ca
Spectral Mode Interference Filter Interference Filter
Detector Photocell
Direct Read Yes Yes
Repeatability ≤3% ≤3%
Linearity
Parameter comparison
Spectral method Interference filter
Display screen LED LCD screen
Sample aspiration rate <6mL/min
Stability Within 15 seconds, ≤ 3%; within 5 minutes, ≤ 5%
Response time <8s
Detection Limit K≤0.004mmol/L;Na≤0.008mmol/L
Repeatability ≤ 3% Relative standard deviation ≤ 3%
Linear error K≤0.005mmol/L;Na≤0.03mmol/L K≤0.005mmol/L、Na≤0.03mmol/L
Measurement range K:0~100ug/ml;Na:0~160ug/ml
Filter Transmittance Characteristics - Peak Wavelength Error Potassium: The absolute value of the peak wavelength error ≤ 7 nm; Sodium: The absolute value of the peak wavelength error ≤ 5 nm.
Filter transmittance characteristics - half-width Potassium: half-width ≤ 15 nm; Sodium: half-width ≤ 15 nm
Transmission characteristics of filters - Background transmittance Potassium: background transmittance ≤ 0.5%; Sodium: background transmittance ≤ 0.5%