Product Name hzaoc GDJW-225D low/high temperature test Chamber hzaoc GDJW-100D low/high temperature test Chamber
Product Price $7771.00 $7125.00
Details Details
Model comparison
Test Type High/Low Temp High/Low Temp
Min-TEMP -60℃ -60℃
Max-TEMP 150℃ 150℃
working volume 101-225L 51-100L
Temp Fluctuation ±0.5℃ ±0.5℃
Temp Uniformity ≤2.0℃ ≤2.0℃
Temp Deviation ±2.0℃ ±2.0℃
Temp Change Rate 2-3℃/min 2-3℃/min
Cooling Method
Parameter comparison
Temperature Range -60℃~150℃ -60℃~150℃
heating rate 1.0~3.0℃/min 1.0~3.0℃/min
cooling rate 0.7~1.0℃/min 0.7~1.0℃/min
Temperature Uniformity ≤ +/- 2.0℃ (when no load) ≤ +/- 2.0℃ (when no load)
Temperature Fluctuation ≤ +/- 0.5 ℃ (when no load) ≤ +/- 0.5 ℃ (when no load)
Temperature deviation ≤±2℃ ≤±2℃
Interior Volume 225L 100L
Inner Chamber dimensions 500*600*750mm(D*W*H) 400*500*500mm(D*W*H)
Dimension 1080*1120*1700mm 1030*1020*1470mm
Time setting Range 0~ 999 hours 0~ 999 hours
Power supply Voltage 380V±10% 380V±10%
Total Equipment Power 3Kw~13.5Kw 3Kw~13.5Kw
power requirements AC380 (+/- 10%) V/50HZ three-phase five-wire system AC380 (+/- 10%) V/50HZ three-phase five-wire system
Inside material of the box 1.0mmSUS304B stainless steel plate 1.0mmSUS304B stainless steel plate
Box outside material 1.0Mm Cold rolled steel plate 1.0Mm Cold rolled steel plate
thermal insulation material Ultra-fine Glass Insulation Cotton Ultra-fine Glass Insulation Cotton
Observation window material Multilayer conductive film Tempered Hollow Glass Multilayer conductive film Tempered Hollow Glass
Observation window Size 200*300mm 200*300mm
test hole Φ50mm Φ50mm
Air supply method Upper air supply and lower air return method Upper air supply and lower air return method
Heating device Porcelain frame nickel-chromium wire electric heating device Porcelain frame nickel-chromium wire electric heating device
Ambient Temperature 5 ℃~ + 28 ℃ (average temperature within 24 hours ≤ 28 ℃) 5 ℃~ + 28 ℃ (average temperature within 24 hours ≤ 28 ℃)
Environmental humidity ≤85% ≤85%
Water source requirements Deionized water or purified water Deionized water or purified water
Meet the standard GB/T2423.1-2001、GB/T2423.2-2001、GB/T10592-2008、GB/T11158-89、GB/T10589-89、IEC68-2-1 GB/T2423.1-2001、GB/T2423.2-2001、GB/T10592-2008、GB/T11158-89、GB/T10589-89、IEC68-2-1