$ 15210.00Ex works
Using multi-channel mechanical compression refrigeration system, energy saving up to 80% of congeneric pRoducts; with 5 ℃/min rapid temperature change rate; through forced air duct system to ensure Temperature Uniformity ≤ 2 ℃; equipped with remote monitoring and fault self-diagnosis function.
Category : High Low Temperature Test Chamber
Brand : HUDA
SKU : NB041429
Service :The delivery and service By NBCLab.Com
HUDA QD-E708-512K60-X5 rapid temperature change test chamber is a professional equipment used to determine the adaptability of electronic products, automotive parts, materials, etc. under conditions of high humidity, low humidity, constant heat and humidity, high and low temperature rapid alternating cycle, heat and humidity of surface condensation. The testing machine adopts a reasonable structure and a stable and reliable Control mode, with beautiful appearance, easy operation, safety, high temperature and humidity control accuracy, suitable for rapid temperature change test, constant temperature and humidity test and alternating humidity cabinet test.
1. Adopt multi-channel, environmentally friendly and energy-saving (low noise) mechanical compression refrigeration system, use environmentally friendly materials, save energy through two-stage multi-channel main bypass system and frequency conversion technology implementation, energy consumption only accounts for 80% of congeneric products, and reduce vibration and noise. 2. The self-developed multi-channel and multi-stage throttling system automatically controls the capillary tubing throttling through the intelligent control system to ensure adjustable flow and uniform temperature field. 3, the application of broadband compression equilibrate temperature control system, the use of temperature reconciling chamber, static pressure chamber and centrifugal fan blade forced air duct system, combined with fast convergence of the PID adaptive algorithm, to achieve rapid temperature control and uniformity. 4. Realize remote visual control, configure man-machine management interface and touch mode automation control system, connect to remote computer through RS485 communication interface, graphical interface for easy management, automatic trouble prompt and provide solutions.
The equipment achieves rapid temperature change and uniform distribution through multi-channel mechanical compression refrigeration system and multi-stage throttling system, combined with intelligent control to automatically adjust the refrigeration output and capillary tubing throttling. The broadband compression equilibrate temperature regulation system is adopted, and the temperature reconciliation chamber, static pressure chamber and centrifugal fan blade forced air duct system are used to ensure full exchange of hot and cold and uniform air supply to the studio. Fast convergence temperature control is achieved through PID adaptive algorithm. The humidification system adopts electric boiler humidification and Compressor dehumidification. P.I.D controls the humidification amount to ensure the accuracy of temperature and humidity.
Electronics, automotive parts, materials
GB2423.1-2008/IEC6008-2-1-2007、GB/T 2423.2-2008、GJB150.3A-2009、GJB150.4A-2009、GB2423.22-2008、GBT 2424.5-2006、GB/T 10589-2008、GB/T2423.3-2008(IEC68-2-3)、GJB150.9A-2009、G/BT 2423.4-2008/IEC6008-2-30:2005、GB/T5170.18-2005、GB/T10586-2006、GBT 2424.6-2006、GBT 2424.7-2006
1. Connect the power supply to ensure that the voltage is AC380V three-phase four-wire system + protective grounding, the voltage fluctuation range is 10%, and the frequency is 50 +/- 0.5Hz. 2. Set the temperature and humidity parameters through the control panel, and you can choose the program mode or the fixed value mode to operate. 3. Start the device, and the system automatically executes the program to monitor the actual temperature and humidity. 4. Use a remote computer to remotely monitor and troubleshoot through the RS485 communication interface. 5. After the test is completed, turn off the power of the equipment.
• The installation site needs to be flat, well ventilated, free of flammable, explosive, corrosive gases and dust, and there is no strong electromagnetic radiation source nearby. • The ground load-bearing capacity is not less than 600kg/m ², and the maintenance space around the equipment is not less than 400mm (A and B directions) and 1000mm (C directions). • The water quality of the water supply must meet the second-level water use standard of the National Analytical Laboratory, the solid impurity content is less than 1mg/L, and the water supply pressure is 0.1 to 0.4Mpa. • Working environment temperature 5~ 30 ℃, humidity ≤ 85% R.H. • The equipment has a variety of protection devices, such as Compressor overheating, overcurrent, overpressure protection, humidification system dry burning protection, power leakage protection, etc., which need to be checked regularly. • Use the whole machine freight method during transportation to avoid violent vibration.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.