Overview
DRP-9052 electric thermostatic incubator consists of five parts: box body, liner (studio), heating system, temperature control system and air circulation system. The box body is sprayed with high-quality steel plate, the liner is made of mirror stainless steel plate with semi-circular arc inner angle for easy cleaning, the liner shelf can be adjusted high and low, filled with fiberglass heat insulation, the inner door is tempered Glass for easy observation, and the silicone sealing ring is used between the studio and the Glass door to ensure high sealing.
Features
1. Microcomputer controller with over-temperature alarm function, stable and reliable use
2. Mirror stainless steel liner
3. Using silicone sealing strip, Glass door is installed on the inner layer of the outer door for easy observation; when the box door is opened, the breeze circulation heating automatically stops, and there is no overshoot
Principle
The product provides heat through a heating system, the temperature control system uses a microprocessor with fuzzy PID control technology and a High Accuracy Platinum Resistance Sensor to achieve precise temperature control, and the air circulation system ensures uniform temperature distribution and maintains a thermostatic environment in the studio.
Applications
Biology, agriculture, scientific research
Steps
1. Connect the AC220V power supply
2. Set temperature and time parameters through the LCD controller
3. Place foster items on adjustable brackets
4. Close the box door to start the device
5. Remove the item and turn off the power after the foster is complete
Notice
• The environment temperature should be kept at 5~ 30 ° C
• relative humidity not more than 80%
• The altitude does not exceed 2000m
• Power supply must meet AC220V +/- 10%, 50HZ +/- 1Hz requirements
• When the box door is opened, the circulating heating automatically stops to avoid opening the door for a long time
• Regularly clean mirror stainless steel liner to maintain hygiene