Overview
The YH1200 haze meter (general purpose) can easily implement the ASTM D1003 non-compensation method and ISO 13468 compensation method for full light transmitance and haze testing. Its open sample compartment supports vertical and horizontal testing to accommodate more samples under test. Using PD array detectors to meet the CIE V (lambda) 2 degree visual Response, compensation method can achieve High Accuracy and repeability light transmitter and haze data measurement.
Features
1. Dual-standard ISO & ASTM to meet the testing standards of different users
2, easy to operate, faster and more accurate measurement, large size touch screen, PD array detector, USB data output device
3. Dynamic measurement, independent Illuminant detector and temperature sensor, always monitor Illuminant and environment changes
4. With compensation port, make the measurement data more accurate, meet the ASTM D1003 non-compensation method and ISO 13468 compensation method
5, easy measurement, wide sample adaptation, open measurement area can be vertical, horizontal test
6. Quality management software, providing powerful haze and light transmitance measurement and analysis software, supporting PC-side data management
Principle
The YH1200 hazy meter adopts transmission illumination mode 0/D (parallel light illumination, diffuse reflection reception), uses φ 154mm integrating sphere and 400~ 700nm combined LED Illuminant. PD array detector satisfies CIE V (lambda) 2 degree visual Response, and achieves High Accuracy and repeatability of light transmitter and hazy data through compensation measurement.
Applications
Glass Processing, Plastic Processing, Film, Display Processing, Encasement Industry, Liquid Liquid Analysis
Standards
ASTM D1003/1044、ISO 13468、ISO 14782、GB/T 2410、JJF 1303-2011、CIE 15.2、JIS K7105、JIS K7361、JIS K 7136
Steps
1. Instrument preheating correction
2. Place the sample in the open sample compartment (can be vertical or horizontal)
3. Measurement mode through large Touchscreen
4, start the measurement, about 1.5 seconds to complete the test
5. Output data through USB interface or connect to quality management software
Notice
• Temperature range from 0 to 40 ° C (32 to 104 ° F)
• Storage Temperature range -20~ 50 ° C (-4~ 122 ° F)
• Sample thickness should be less than 170mm
• Power supply with DC 24V, 3A Power Adapter
• Ensure that the instrument is fully preheated and calibrated before measurement