Overview
KESIONOTS KS-OTS-HD1000LB low/high temperature test chamber, also known as environmental test chamber, is suitable for high temperature and low temperature Reliability test of industrial products. The equipment can be used for electronic and electrical, automobile and motorcycle, aerospace, ship weapons, colleges and universities, R & D institutions and other related products. The parts and materials of high temperature and low temperature (alternating) cycle changes in the inspection of various performance indicators for product design, improvement, identification and inspection.
Features
1. Temperature range -60~ 150 ℃
2. Adopt stainless steel studio and high-grade cold plate nano paint outer box
3. Equipped with high temperature resistant high density chlorine-based formic acid vinegar foam insulation material
4. Optional air-cooled or water-cooled dual-stage Compressor refrigeration system
5. Configure multiple protection devices including no fuse switch, Compressor overload protection, etc
6. Adopt Korean or Japanese "OYO" brand controller system
7. Equipped with French "Taikang" brand or German "BITZER" refrigeration system
Principle
This machine realizes high temperature and low temperature environment simulation through precise temperature control system, and promotes the performance test of product materials under temperature change conditions. For example, in the aging test, the deterioration of sulfur rubber is promoted by heating, and the tensile force and elongation change rate before and after heating are calculated. One day of testing at 70 ° C is theoretically equivalent to 6 months of exposure to the atmosphere.
Applications
Electronics and electricians, automobiles and motorcycles, aerospace, ships and weapons, colleges and universities, R & D institutions
Standards
GB10592-89、GB2423.1、GB2423.2、GJB150.3、GJB150.4、IEC、MIL
Steps
1. According to GB 2423.1, GB 2423.2 standards for low temperature, high temperature test
2. Conduct a constant temperature test
3. Set the required temperature parameters through the controller system
4. Observe the performance of the test sample under high temperature and low temperature cycle changes
Notice
Equipment is equipped with multiple protection devices to ensure safe operation
• Power supply voltage compliance before use (1 220VAC +/- 10% 50/60Hz & 3 380VAC +/- 10% 50/60Hz)
• Regularly check the working status of the refrigeration system
• Pay attention to Temperature Uniformity and control accuracy indicators
• Conduct test operations according to standard requirements