Please contact customer service
Temperature range -60 ℃ to 150 ℃, Temperature Uniformity up to +/- 1 ℃, stainless steel inner box and high-efficiency insulation material, support Air-Cooled/Water-Cooled multiple Cooling modes, suitable for PRoduct Research & Development and mass inspection.
Category : High Low Temperature Test Chamber
Brand : kesionots
SKU : NB042711
Service :The delivery and service By NBCLab.Com
KESIONOTS KS-HD80LB low/high temperature test chamber, also known as Environmental Test Chamber, is suitable for high temperature and low temperature Reliability test of industrial products. The equipment is used for high temperature, low temperature (alternating) cycle change conditions for parts and materials of electronic and electrical, automotive and motorcycle, aerospace, marine weapons, colleges and universities, R & D institutions and other related products for performance inspection, serving product design, improvement, identification and inspection, such as aging test.
1. Temperature range is wide, up to -60 ℃ to 150 ℃. 2. Temperature control accuracy, analytical accuracy of +/- 0.1 ℃, control accuracy +/- 1 ℃. 3. The box is made of excellent material, the outer box is high-grade cold-rolled sheet paint, and the inner box is stainless steel. 4. Equipped with a variety of protection devices, such as no fuse switch, Compressor overload protection, refrigerant high voltage protection, over temperature protection, etc. 5. Optional well-known brand controller (Korean "TEMI" or Japanese "OYO") and Compressor ("Taikang" or German BITZER). 6. Provide standard accessories, such as viewing windows, test holes, box lights, Shelf, etc.
The low/high temperature test chamber simulates and precisely controls the high temperature, low temperature or alternating temperature environment in a closed box through the refrigeration system (Air-Cooled/single-stage Compressor or wind, Water-Cooled/double-stage Compressor) and heating system. Control system (such as TEMI or OYO controller) adjusts the cooling and Heating power according to the setting program, so that the temperature of the working room changes according to the requirements, to test the performance and Reliability of the product under extreme temperature conditions.
Electronics and electricians, automobiles and motorcycles, aerospace, ships and weapons, colleges and universities, R & D institutions
GB2423.1、GB2423.2、GJB150.3、GJB150.4、IEC、MIL
1. Place the sample to be tested on the Shelf in the box. 2. Close the door and make sure the sealing is good. 3. Set the required temperature program (high temperature, low temperature or alternating cycle) through the controller. 4. Start the equipment, the test chamber starts to run, and the temperature is raised and cooled according to the set program. 5. During the test, the sample status can be observed through the viewing window. 6. After the test, the equipment is automatically stopped or manually stopped, and the sample is taken out after the temperature in the box returns to a safe range.
• Device power supply needs to match, support 220VAC +/- 10% 50/60Hz or 380VAC +/- 10% 50/60Hz. When placing the sample, ensure that the air circulation in the box is not affected, and the total volume of the sample does not exceed 1/3 of the volume of the studio. • During operation, do not open the door frequently, so as not to affect the Temperature Stability and test results. • The equipment is equipped with a variety of protection devices. If the alarm system is triggered, please check the cause of the failure first and do not force it to run. • Regularly check the chilling down system (such as Water-Cooled type, make sure the chilling down water supply is normal) and Compressor status. • When conducting extreme low temperature (e.g. below -40 ° C) tests, it is necessary to confirm that the equipment is equipped with a corresponding dual-stage Compressor refrigeration system. The cooling rate of the equipment is about 1.0 ° C/min. If a faster rate is required (5-10 ° C/min), it is a special selection condition and needs to be confirmed in advance.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.